The Model T703U Photometric O3 Calibrator is designed to meet the requirements for low level ozone calibrations and audits, which requires stable, repeatable ozone generation at levels far below the capability of standard ozone calibrators.
With NumaView™ premium T Series software
- Large, vivid, and durable color touchscreen display
- Lifetime technical support by phone and email
- All other T Series instrument platform features
- Standard two-year warranty
System
Linearity | ± 1% of full scale |
Precision | 1.0 ppb |
Response Time | < 240 seconds to 95% |
Stability (7 days) | < 1% with photometer feedback < 3% without photometer feedback (CNST or REF) |
Ozone Generator Module
Flow Rate (with internal zero air source) | 1 to 5 LPM adjustable |
Flow Rate (with external zero air source) | 1 to 15 LPM adjustable |
Maximum Output | 5 ppm LPM |
Minimum Output | 15 ppb LPM |
Maximum Concentration | 5 ppm at 1 LPM |
Minimum Concentration | 3 ppb at 5 LPM |
Response Time | < 240 seconds to 98% |
UV Photometer
Range | 0 - 100 ppb to 0 - 10 ppm (selectable) |
Precision | 1.0 ppb |
Linearity | ± 1% of full scale |
Rise/Fall Time | < 20 seconds to 95% (photometer response) |
Response Time | < 180 seconds to 95% (system response) |
Zero Drift | < 1.0 ppb/7 days |
Span Drift | < 1%/24 hours |
Lag Time | < 10 seconds |
Flow Rate | 800 cc/min ±10% |
Electrical Specifications
Power Requirements | 100V-120V, 220V-240V, 50/60 Hz |
Analog Output Ranges (Test Channel) | 10V, 5V, 1V, 0.1V (selectable) |
Communications Specifications
Included I/O | 1 x Ethernet: 10/100Base-T 2 x RS232 (300-115,200 baud) 2 x USB device ports 12 x digital control outputs 12 x digital control inputs 8 x digital status outputs |
Optional I/O | 1 x USB com port 1 x RS485 Multidrop RS232 |
Physical Specifications
Operating Temperature Range | 5 - 40ºC |
Dimensions (H x W x D) | 7” x 17” x 24” (178 x 432 x 609 mm) |
Weight | 35.5 lbs (16.1 kg) with internal zero air pump |
Specifications subject to change without notice.